Thermo Scientific Isotemp -86 C Freezers Assembly, Installation And Operation page 29

Table of Contents

Advertisement

S.
Function
Parameter
No
Code
Cold Alarm
03
0x03
Setpoint (C)
System Bill
of Material
04
0x03
Part
Number
System Bill
of Material
05
0x03
Part
Number
System Bill
of Material
06
0x03
Part
Number
System Bill
of Material
07
0x03
Part
Number
System Bill
of Material
08
0x03
Part
Number
09
Product ID
0x03
Isotemp -86°C Freezers
Address
Modbus
in hexa
Size
Command
decimal
3A 30 31
30 33 30 35
33 43 30 30
053C
2
30 32 42 39
0D 0A
3A 30 31
30 33 30 35
37 30 30 30
570
2
30 32 38 35
0D 0A
3A 30 31
30 33 30 35
37 34 30 30
574
2
30 32 38 31
0D 0A
3A 30 31
30 33 30 35
578
2
37 38 30 30
30 32 37 44
0D 0A
3A 30 31
30 33 30 35
57C
2
37 43 30 30
30 32 37 39
0D 0A
3A 30 31
30 33 30 35
580
2
38 30 30 30
30 32 37 35
0D 0A
3A 30 31
30 33 30 35
594
2
39 38 30 30
30 32 35 44
0D 0A
Relay
Data
Enclosure
Data
Type
Response
3A 30 31
30 33 30 34
0xFF
46 46 46 46
FFF
int
46 43 37 43
C7C
38 32 0D
0A
3A 30 31
30 33 30 34
33 31 33 35
string
33 35 34 34
31 39 0D
0x31
0A
0x35
0x35
3A 30 31
0x44
30 33 30 34
35 32 33 30
string
0x52
34 31 33 30
0x30
30 35 0D
0x41
0A
0x30
0x31
30 31 30 33
0x52
30 34 33 31
string
0x20
35 32 32 30
0x20
32 30 33 35
0D 0A
0x20
0x20
3A 30 31
0x20
30 33 30 34
0x20
32 30 32 30
string
32 30 32 30
0x20
0x20
37 38 0D
0x20
0A
0x20
3A 30 31
30 33 30 34
32 30 32 30
string
32 30 32 30
37 38 0D
0A
3A 31 39
30 33 30 34
0x00
30 30 30 31
01E2
uint
45 32 34 30
40
42 39 0D
0A
Appendix B: Modbus ASCII Parameter Table |
Remarks
Convert the data
value into signed 2's
complement and
divide with 10, which
gives the setpoint.
Ex: Signed 2's
complement of the
0xFFFFFC7C is equal
to –900.
–900/10=–90. So the
CA setpoint is –90C.
To get BOM combine
570,574,578 data.
Ex. 155DR0A01R
All the remaining are
spaces.
All the remaining are
spaces.
123456 would be
encoded as
0x01E240.
25

Advertisement

Table of Contents
loading

Table of Contents