Thermo Scientific Isotemp -86 C Freezers Assembly, Installation And Operation page 27

Table of Contents

Advertisement

Alarm Summary
Alarm Message
System Battery
Low Voltage Alarm
Bus Battery
Low Voltage Alarm
Stuck Button Alarm
System Battery
Failure Alarm
BUS Probe
Failure Alarm
BUS Battery
Failure Alarm
Wrong Power Alarm
Isotemp -86°C Freezers
Remote Alarm
Warning Icon
Event
Wrench
Yes
Wrench
Yes
Wrench
Yes
Wrench
Yes
Wrench
Yes
Wrench
Yes
Wrench
Yes
Description
System Battery voltage is too low. If error persists, the battery
may need to be replaced.
Display will intermittently show "E07".
BUS Battery voltage is too low. If error persists, the battery
may need to be replaced.
Display will intermittently show "E06".
A button has been pressed for more than 5 minutes.
Display will intermittently show "E10".
System Battery disconnected or failed.
Display will intermittently show "E12".
BUS cannot detect temperature. BUS will continuously inject.
Contact Customer service.
Display will intermittently show "E13".
BUS battery disconnected or failed.
Display will intermittently show "E14".
The unit has detected the Wrong Power connected. Verify the
proper voltage.
Display will intermittently show "EA1".
Appendix A: Alarm Summary |