Thermo Scientific Isotemp -86 C Freezers Assembly, Installation And Operation page 22

Table of Contents

Advertisement

Problem
Unit tripping the circuit breaker.
Excessive frost build-up
around perimeter of door.
Unit is over cooling.
Unit compressors run
continuously.
Cabinet temperature reached
an alarm
condition, but suitable alarm is
not activated.
Problem with temperature
validation/calibration.
18 |
Troubleshooting Guide
Cause
Shared power source.
Unit plugged into wrong
power outlet.
Unit not grounded.
Use of extended cords.
Icing/Frost due to high
relative humidity.
Excessive and prolonged
door openings.
Gasket damage.
Set points may have
changed.
Temperature offset may
have changed.
Unknown.
Freezer set point is low.
Frost build up.
Dirty condenser.
Gasket damage.
Alarm setpoints may be
changed.
Cabinet temperature
displayed doesn't match
with actual temperature.
Solution
Never connect unit to overloaded power source. Always
use a dedicated (separate) circuit.
Plug the unit into proper power source to deliver correct
voltage.
Your unit must be properly grounded in conformity with
national and local electrical codes. Troubleshooting
procedures involving live voltage is dangerous and if done
improperly can result in injury and/or death. This
troubleshooting should be performed by trained personnel
only.
Do not use an extension cord. Make sure the unit supplied
power cord is plugged directly into power outlet.
Check if the location meets requirements. Maximum relative
humidity 60% for temperatures within 15°C to 32°C (59°F
to 90°F).
Occasionally scrape the ice on the outer door.
Be careful not to puncture the rubber gasket.
Avoid opening door for a prolonged time.
Check for punctures or tears on gasket. If replacement is
necessary, call service. Refer Gasket Maintenance in user
manual.
Adjust the setpoint to run at desired setpoint under settings.
Try adjusting the offset. Temperature offset can be set by
accessing the settings menu via the settings button.
Try re-starting the unit. If this doesn't help call service.
Check whether the setpoint is in operating range. Change
the setpoint if necessary.
Defrost the unit. Refer Defrosting the Freezer in user
manual.
Clean the condenser and condenser filter.
Check for punctures or tears on gasket. If replacement is
necessary, call service. Refer Gasket Maintenance in user
manual.
Check the present setpoints for temperature alarm
conditions. Change the setpoints if required.
Customers performing on-site temperature calibration may
observe as much as a 2°C variation when an external probe
is placed next to the freezer control probe. This variation is
normal due to optimisation of the control system to ensure
temperature uniformity throughout the cabinet.