Thermo Scientific Isotemp -86 C Freezers Assembly, Installation And Operation page 32

Table of Contents

Advertisement

S.
Function
Parameter
No
Code
16
TC5
0x03
17
TC6
0x03
18
TC7
0x03
28 |
Appendix B: Modbus ASCII Parameter Table
Address
Modbus
in hexa
Size
Command
decimal
3A 30 31
30 33 30 34
04D8
2
44 38 30 30
30 32 31 45
0D 0A
3A 30 31
30 33 30 34
04DC
2
44 43 30 30
30 32 31 41
0D 0A
3A 30 31
30 33 30 34
04E0
2
45 30 30 30
30 32 31 36
0D 0A
Relay
Data
Enclosure
Data
Type
Response
3A 30 31
30 33 30 34
0x42
34 32 31 37
17D0
Float
44 30 30 31
01
43 45 0D
0A
3A 30 31
30 33 30 34
0x42
34 32 31 37
17D0
Float
44 30 30 31
01
43 45 0D
0A
3A 30 31
30 33 30 34
0x42
34 32 31 37
17D0
Float
44 30 30 31
01
43 45 0D
0A
Remarks
Convert the Float to
decimal with below
steps:
1.Open the link
https://www.h-
schmidt.net/
FloatConverter/
IEEE754.html
2.Paste the data in
"Hexadecimal
Representation" and
press Enter.
3.The temp value is
shown in "Decimal
representation"
Convert the Float to
decimal with below
steps:
1.Open the link
https://www.h-
schmidt.net/
FloatConverter/
IEEE754.html
2.Paste the data in
"Hexadecimal
Representation" and
press Enter.
3.The temp value is
shown in "Decimal
representation"
Convert the Float to
decimal with below
steps:
1.Open the link
https://www.h-
schmidt.net/
FloatConverter/
IEEE754.html
2.Paste the data in
"Hexadecimal
Representation" and
press Enter.
3.The temp value is
shown in "Decimal
representation"