Appendix B: Modbus ASCII
Parameter Table
Protocol
MODBUS ASCII
Baud Rate
2400bps to 57.6Kbps
Data Bits
7
Stop Bits
1
Parity
Even
Flow Control
None
Address
0 to 255
S.
Function
Parameter
No
Code
Cabinet
01
0x03
Setpoint (C)
Warm Alarm
02
0x03
Setpoint (C)
24 |

Appendix B: Modbus ASCII Parameter Table

Address
Modbus
in hexa
Size
Command
decimal
3A 30 31
30 33 30 35
33 30 30 30
530
2
30 32 43 35
0D 0A
3A 30 31
30 33 30 35
538
2
33 38 30 30
30 32 42 44
0D 0A
Relay
Data
Enclosure
Data
Type
Response
3A 30 31
30 33 30 34
0xFF
46 46 46 46
FFF
int
46 43 45 30
CE0
31 45 0D
0A
3A 30 31
30 33 30 34
0xFF
46 46 46 46
FFF
int
46 44 34 34
D44
42 39 0D
0A
Remarks
Convert the data
value into signed 2's
complement and
divide with 10, which
gives the setpoint.
Ex: Signed 2's
complement of the
0xFFFFFCE0 is equal
to -800.
–800/10= –80.
So the setpoint is
–80C.
Convert the data
value into signed 2's
complement and
divide with 10, which
gives the setpoint.
Ex: Signed 2's
complement of the
0xFFFFFD44 is equal
to –700.
–700/10= –70.
So the WA setpoint is
–70 C.
Isotemp -86°C Freezers