Mini Incubator Operation Manual
Model: MINI INCUBATOR
NB-201M

Advertisement

loading

Summary of Contents for N-BIOTEK NB-201M

  • Page 1 Mini Incubator Operation Manual Model: MINI INCUBATOR NB-201M...
  • Page 2: Table Of Contents

    CONTENTS PAGE WARRANTY GENERAL INFORMATION ON PRECAUTION TRANSPORATION, STOARGE AND LOCATION OF INSTALLATION PREREQUISITE AND CONFIGURATION FEATURE AND SPECIFICATION CONTROL PANEL OPERATION Page 2 / 15...
  • Page 3: Warranty

    NBIOTEK, whichever is sooner as indicated in above table. This period is proved by serial number. 3. N-BIOTEK will not be responsible of free warranty service for the faulty caused by user’s improper operation, excessive use, use of incorrect voltage & frequency, storage in wrong environment mentioned in Manual.
  • Page 4: General Information On Precaution

    ⚫ General Information on Precaution Precaution is to prevent the possible accident or danger during operation. So, you must keep it. Precaution is divided into caution and warning. And, each of them has following meanings. Other marks.. 1. Precaution for using the power cable Do not make the power plug be pressed by back of the product.
  • Page 5 2. Precaution for ground connection Please ground before use the product, if you don't ground, you might get an electrocution when malfunction or an electric leakage occurs. At the place where you can't ground, * Please buy the equipment to prevent any electrical leakage. * An electric shock, an electric leakage and a fire can be occurred without an electric leakage breaker.
  • Page 6: Transporation, Stoarge And Location Of Installation

    Please leave the product power off when it is not in use. (It is to prevent an eclectic leakage.) Open and close the door softly and please use a door knob. (A heavy shock can damage the product and breakdown the operating part.) Do not detach the built-in lamp and electrical devices.
  • Page 7 3. Location of installation and ambient conditions Do not put flammable substances near the product. (It can cause a fire) When you install the product, you have to put the distance of at least 30cm from the wall. To completely separate the unit from the power supply, power plug must be disconnected.
  • Page 8: Prerequisite And Configuration

    ⚫ Prerequisite and Configuration 1. Prerequisite Inspection of Boxes When you have received the instrument which is well boxed, inspect the box carefully to check any transit damage. Please report any damage to the carrier or to your local NBIOTEK distributor immediately. Location The incubator is designed to operate at temperature 5℃...
  • Page 9 2. Configuration. <Figure #1> Front side Handle for carrying Control panel Door Handle Finger Supporter for door close <Figure #2> Inner side and rear side Door Sealing Gasket Stainless steel panel Stainless Steel shelf Circulation fan Water Container (Only include in CO₂ Model, NB203M) Page 9 / 15...
  • Page 10 <Figure #3> Rear of Incubator Serial Label Power Switch Power Input Receptacle / DC12V Car Jack Receptacle Air Circulation Cooling Fan Foot of Incubator with non-slip rubber <Figure #4> Basic Parts included in the package 1. Stainless Steel Shelf (2) 2.
  • Page 11: Feature And Specification

    ⚫ Feature and Specification 1. Feature 1. Light weight, easy to carry concept of 16liter potable mini incubator. 2. Compact size for the use inside workstation or clean bench. 3. Precise temperature control by Peltier. (ambient -8’C, lowest temp. is 15’C at 24’C RT) 4.
  • Page 12: Control Panel

    ⚫ Control Panel 1. HEATING INDICATION LAMP 2. COOLINGING INDICATION LAMP 3. Temperature LED display 4. Temperature setting button 5. Adjustment button Page 12 / 15...
  • Page 13: Operation

    ⚫ Operation 1. Place and install the product. Install the product at the desired place and check the level in all directions (side by side, front to back and ground). 2. Connect the power Prior to connect the power plug, make sure that the POWER S/W is off 3.
  • Page 14 a. Press and hold “CAL/SET” for 10 seconds. Then, LED will be flickering as below. Channel 1 is at chamber’s Main Temp calibration stage. Press UP (▲) as much as difference from measured value by precise analyzer if it is higher. Press DOWN (▼) as much as difference from measured value by precise analyzer if it is lower.
  • Page 15 6. Alarm Turn alarm system On by pressing alarm button for about 5 seconds. When switching alarm on or off, a short alarm will come out. When alarm is on, LED LAMP off. When Alarm system is off, LED LAMP of alarm is lighting to indicate that alarm system is disarmed.