  • Page 1 OPERATING MANUAL ORBITAL SHAKER MODEL : NB-101S...
  • Page 2 The power outlet must be only for this product. (Using various products simultaneously can cause a fire) Clean the power plug with a dry towel and connect it properly. (Foreign substances or unsafe connection can cause a fire.) NB-101S(Ver. 2014) 2/13...
  • Page 3: Precaution For Installation

    When you install the product, you have to put the distance of more than 20cm from the wall. Install at a flat and solid ground. (If the ground is not flat, it can cause a vibration of the product.) NB-101S(Ver. 2014) 3/13...
  • Page 4: Precaution For Use

    Do not wash the product with excessive quantity of water, thinner, benzene and Petroleum. (It can cause an electric leakage, and malfunction or damage on the surface.) When you don't use the product or clean it, please pull out the power plug. (It is to prevent an eclectic leakage.) NB-101S(Ver. 2014) 4/13...
  • Page 5 If you don't have the outlet for AC 220V, then bury it under the ground after connecting the ground line to copper plate. * No ground connection can bring an electrocution, an electric leakage and a Fire. NB-101S(Ver. 2014) 5/13...
  • Page 6: Feature & Specification

    Magnetic induction drive & Brush less DC motor provides less vibration and maintenance. ● Artificial intelligence system which maintains precise speed & adjust time. ● LED indicates temperature, speed, time and state of power failure. ● No heat generated during operation. ● Easy to change accessory platform. 2.2 Specification NB-101S(Ver. 2014) 6/13...
  • Page 7: Power Switch

    CONTROL PANEL 2.4 Control panel DISPLAY For displaying RPM and TIME RPM/TIME Key for selecting RPM and TIME Key for saving desired RPM and TIME UP, DOWN Key for adjusting RPM and TIME value START/STOP START / STOP NB-101S(Ver. 2014) 7/13...
  • Page 8: Operation

    At first, you can set the hour using UP(▲), DOWN(▼) key and press “SET” key. You can set the minute using UP(▲), DOWN(▼) key after set the hour. Press “SET” key again after set time (hour and minute), then LED display as “t SAVE” as below. NB-101S(Ver. 2014) 8/13...
  • Page 9 When you set the time like above and operate instrument, -END- is displayed after a minutes. But shaking is well running without stop till reaching desired time. After pushing start button under time set-up, the instrument would be stopped with displaying –END- when the time is reached at set time. NB-101S(Ver. 2014) 9/13...
  • Page 10: Accessory Description

    250ml Flask Holder x 9ea 500ml Flask Holder x 5ea with 101S-PL with 101S-PL with 101S-PL 101S-H100 101M-M Tube rack + 101S-PL 1000ml Flask Holder x 4ea Microplate Rack<6ea> With 101S-PL With 101S-PL 101S-2S 2 Stairs (4 Supporters with 101S-PL) NB-101S(Ver. 2014) 10/13...
  • Page 11 - Momentary noise symptom. - When the platform is being moved faster than RPM which you set. (This message may occur from 200 RPM) - Solution : These symptoms will be recovered automatically within 3 min or restart it. NB-101S(Ver. 2014) 11/13...
  • Page 12 - When the platform is being moved faster than RPM which 5) This message may be marked as to - The electricity is off during operation. - Solution : "n"(new record) will be changed to "r"(RPM) if you push "START" button. NB-101S(Ver. 2014) 12/13...
  • Page 13: Warranty

    N-BIOTEK, whichever is sooner as indicated in above table. This period is proved by serial no. 3. N-BIOTEK will not be responsible of free warranty service for the faulty caused by user’s improper operation, excessive use, use of incorrect voltage & frequency, storage in wrong environment mentioned in Manual.