Example: 4 X 4 Matrix Switching; Example: Dual 4 X 2 Matrix Switching - HP E1361A 4X4 User Manual

Relay matrix module
Table of Contents

Advertisement

Example: 4 X 4
Matrix Switching
Example: Dual
4 X 2 Matrix
Switching
26 Using the HP E1361A Relay Matrix
All manuals and user guides at all-guides.com
This example closes channel 30 (relay 30) of a relay matrix in 4 X 4
configuration to connect the user input to row 3 with the user input to
column 0. See Figure 3-1 for typical user connections. To close channel 30,
execute:
CLOS (@130)
To open channel 30 (relay 30), use
Figure 3-1. Example: 4 X 4 Matrix Switching
This example closes relays 00 and 02 (channels 00 and 02) for a relay
matrix in dual 4 X 2 configuration. Closing relay 00 connects the user input
to Bank A Row 0 with the user input to Column 0. Closing relay 02
connects the user input to Bank B Row 0 with the user input to Column 2.
See Figure 3-2 for typical user connections (note that JM1-JM8 must be
removed for dual 4 X 2 matrix configuration). To close channels 00 and 02
(relays 00 and 02), execute:
CLOS (@100,102)
To open channels 00 and 02, use
Close relay 30 (1 is card number)
OPEN (@130)
.
Close channel 00 and 02 relays. 1
is the card number and 00 and 02
are the channel numbers.
.
OPEN (@100,102)