Subsystem Syntax

:CONTinuous

Parameters
Comments
Example
46 Command Reference
All manuals and user guides at all-guides.com
The
command subsystem selects continuous scanning cycles and starts the

INITiate

scanning cycle.
INITiate
mode
:CONTinuous <
:CONTinuous?
[:IMMediate]
mode
INITiate:CONTinuous <
the switchbox.
Parameter Name
mode
Continuous Scanning Operation: Continuous scanning is enabled with the
INITiate:CONTinuous ON
command closes the first channel in the channel list. Each
INITiate:IMMediate
trigger from a trigger source selected by the
advances the scan through the channel list. A trigger at the end of the channel list
closes the first channel in the channel list and the scan cycle repeats.
Non-Continuous Scanning Operation: Non-continuous scanning is enabled
with the
INITiate:CONTinuous OFF
Sending the
INITiate:IMMediate
list. Each trigger from a trigger source selected by the
command advances the scan through the channel list. At the end of the scanning
cycle, the last channel in the channel list is closed.
Stopping Continuous Scan: See the
Related Commands:
ABORt
*RST Condition:
INITiate:CONTinuous OFF
Enabling Continuous Scanning
This example enables continuous scanning of channels 00 through 03 of a
single-module switchbox. Since
an interface clear command (such as CLEAR) to stop the scan.
INIT:CONT ON
SCAN (@100:103)
INIT
>
enables or disables continuous scanning cycles for
>
Parameter Type
boolean
or
INITiate:CONTinuous 1
TRIGger:SOURce
or
INITiate:CONTinuous 0
command closes the first channel in the channel
ABORt
command.
,
,
ARM:COUNt
TRIGger
TRIGger:SOURce IMMediate
Enable continuous scanning
Defines channel list
Start scan cycle, close channel 00
INITiate
Range of Values
ON | OFF | 1 | 0
command. Sending the
command
command.
TRIGger:SOURce
,
TRIGger:SOURce
(default) is set, use
Chapter 5