Icare HOME2 Quick Manual page 17

Hide thumbs Also See for HOME2:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
va
Icon
Display text
Sound
3 long
TOO FAR
5 short
TOO NEAR
2 short
INCORRECT
ALIGNMENT
2 short
REPEAT
DETECTION
2 short
ERROR
Problem
The probe is too far away from the
eye.
The probe is too close to the eye.
The probe was not perpendicular to
the cornea or the probe hit an
eyelid or eyelashes.
There was a problem with the
movement of the probe.
The sensor could not identify the
measured eye (right or left).
u
Solution
Press the Measure button
once to acknowledge the error.
Shorten the forehead and cheek supports.
Press the Measure button
once to acknowledge the error.
Lengthen the forehead and cheek supports.
Press the Measure button
once to acknowledge the error.
Position the tonometer so that the probe points perpendicu-
larly to the center of your eye. Keep your eye properly open.
Press the Measure button
once to acknowledge the error.
Repeat the measurement.
Press
, then press
until the correct eye (right or left) is
shown on the display. Press
, or press the return button to
cancel the measurement.
17