Power Requirements; Environmental Conditions; Physical Characteristics - Keysight Technologies N9029BV Product Notes

Mini signal analyzer frequency extension module
Table of Contents

Advertisement

Power Requirements

• Voltage and/or range (V): 100-240 Vac (AC input for AC/DC adapter), 9 Vdc (input
for module)
• Frequency and/or range (Hz): 50-60 Hz (AC input for AC/DC adapter)
• Power in Watts, VA or Current (A): 1.2 A max (input for AC/DC adapter), 4 A (input
for module)
The instrument can operate with mains supply voltage fluctuations up to
± 10% of the nominal voltage.
Install the instrument so that the detachable power cord is readily identifiable and is
easily reached by the operator. The detachable power cord is the instrument
disconnecting device. It disconnects the mains circuits from the mains supply before
other parts of the instrument. Alternatively, an externally installed switch or circuit
breaker (which is readily identifiable and is easily reached by the operator) may be
used as a disconnecting device.
An external attenuator is recommended when using N9029BV Mini-SAX with
the EXM option to avoid exceeding the damage input LO power (16 dBm).

Environmental Conditions

• For indoor use only
This product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2.
Be aware of the power level applied to the waveguide input. Non-warrantied
damage to the internal mixer can occur if the power level is at or above the
damage level specified in the VDI User's Guide for "RF Input."
Before using your SAX module, please read all documentation from VDI
(included on the USB drive).

Physical Characteristics

Table 1
Dimensions
Height
Width
Length
4
1.50 inches
5.25 inches
3.50 inches
Keysight N9029-90038