What to do if...
Sensor symbol flashes
Display
Meter does not react to
keystroke
You want to know which
software version is in
the meter or IDS sensor
109
6
What to do if...
6.1
General information
Cause
– Calibration interval expired
Cause
– Batteries almost empty
Cause
– Operating condition undefined
or EMC load unallowed
Cause
– E. g., a question by the service
department
Remedy
– Recalibrate the measuring
system
Remedy
– Charge the batteries (see
section 3.3.2 C
ONNECTING
THE POWER PACK
)
THE BATTERIES
– Replace the batteries (see
section 5.1 M
AINTENANCE
Remedy
– Processor reset:
Press the <ENTER> and
<On/Off> key simultane-
ously
Remedy
– Switch on the meter.
– Open the menu,
_
<ENTER
> / Storage &
config / System / Service
information. The instrument
data are displayed.
or
– Connect the sensor.
Press softkey [Info]/[More].
The sensor data are dis-
played (see section 4.5)
ba75864y18
Multi 3420
/
CHARGING
)