Interrupting a program

A program can only be interrupted
or canceled in operating levels B
and D.
Only interrupt a program if the door
must be opened for urgent reasons,
e.g., if items are moving around too
much, or the cleaning performance
needs to be checked.
If a program with a defined process
requirement is interrupted then
continued, make sure to check the
display upon program completion. If
the display reads
parameter not achieved
was opened after process
parameter monitoring began, and so
the defined requirements were not
achieved. If necessary, repeat the
program.
In operating levels B or D
^ Press the +C button.
At water temperatures below 104°F
(40 °C), the following message appears
in the display:
Program Cancel (OK)
or Continue (Clear)?
At water temperatures above 104°F
(40 °C):
Cancel program (OK)
^ Open door.
Process
, the door
,
Caution! Items in the machine
may be hot. There is the danger of
burning or scalding.
^ Arrange items securely.
^ Carefully close door.
,
If there is a large volume of hot
water in the wash cabinet when a
program is interrupted, and the lab
washer door is closed too quickly,
hot water can leak out.
There is the risk of scalding and
burning.
If, at the time of program interruption,
the water temperature was below 104°F
(40 °C):
^ Press the +C button.
The program continues.
If, at the time of program interruption,
the water temperature was above 104°F
(40 °C):
^ Press the OK button.
The program is canceled.
The display reads:
Program canceled Water drainage
Once the water has drained away the
program can be restarted.