Use; Loading The Machine; Process Validation - Miele PG 8536 AD LFM SST Operating Instructions Manual

Laboratory glassware washer
Table of Contents

Advertisement

The lab washer can be fitted with a
variety of mobile units, baskets and
inserts, depending on the cleaning
application desired.

Loading the machine

Removing excess soils
^ Empty all containers before loading
them into the machine.
,
Ensure that no acid or solvent
residues, especially hydrochloric
acid or chlorides get into the wash
cabinet.
Check before cleaning
– Are the items to clean correctly
sorted and connected?
– Are the spray arms clean and can
they rotate freely?
– Is the filter combination free of
coarse soiling?
– Is the adapter to the water inlet in
place?
– Are the spray arms correctly
connected?
– Are the detergent and neutralizer
containers full?
Check after cleaning
– Check the cleaning results visually.
– Are the jets and connections
securely connected with the mobile
unit / inserts?

Process validation

As a rule, it is the responsibility of the
user to ensure that items cleaned in the
machine meet the required standards.