Troubleshooting - Renishaw QUANTiC RTLC40-S Installation Manual

Incremental linear encoder system
Hide thumbs Also See for QUANTiC RTLC40-S:
Table of Contents

Advertisement

Troubleshooting

Fault
LED on the readhead is blank
LED on the readhead is flashing red
during installation mode
Unable to get a green LED over the
complete axis length
Can't initiate the calibration routine
22
QUANTiC
RKLC40-S incremental linear encoder system
Cause
There is no power to the readhead
The signal strength is < 50%
System run-out is not
within specification
Signal size is < 70%
Ensure 5 V is supplied at the readhead
For cable variants, check the correct wiring of the connector
If using the analogue variant of QUANTiC with the ADTi, ensure the appropriate adaptor cables are connected
Check the readhead optical window and scale are clean and free from contamination
Restore factory defaults (see page 21) and check the alignment of the readhead. In particular:
Rideheight
Yaw
Offset
Ensure the correct scale and readhead combination
Use a DTi gauge and check the run-out is within specifications
Restore factory defaults (see page 21)
Realign the readhead to obtain a green flashing LED at the mid-point of the run-out
Recalibrate the system (see page 20)
Realign the readhead to obtain a green flashing LED
Possible solutions