Troubleshooting
Error
E-41
|
E-56
E-60
54
Symptom
Rotor stops without
deceleration to standstill.
Instrument cannot be
operated.
Rotor stops with decelera-
tion.
Possible causes and corrective measures
Internal program error.
Switch the instrument off and on again.
If the error persists, contact service representative.
Insufficient temperature in the refrigeration unit.
1. Stop the centrifugation run.
2. Open the centrifuge lid, defrost the chamber
Never touch the chamber directly with
your hands – you may freeze up!
3. After approx. 60 minutes you can restart the instrument.
Observe the maximum permissible environmental
temperature!
4. If a strong ice sheet is present in the internal chamber,
be sure to remove all condensate after defrosting
If the error persists, contact service representative.