Correct operation when filling the sample tubes and
closing the rotor lid are prerequisites for aerosol bio-
containment.
Prior to each application, the seals in
the rotors have to be checked for
correct fit, abrasion or damage and
have to be slightly greased.
Replace damaged seals immediately!
Use only the special lubricant
7600 3500 to grease the seals!
A set of spare seals is supplied with the rotor and can
be ordered separately as spare seal set 75003405.
The following steps have to be carried out:
Lubricate the seals before inserting them
(lubricant order no. 76003500)
Press the smaller V-seal into the groove of the
rotor collar (a).
Insert the seal (C-profile) in the outer groove of the
rotor body (b).
Pay attention after loading the rotor to
safe closing of the rotor cover!
Replace damaged or dulled rotor
lids promptly!
Handling the rotor
25