Error
„bAL"
E-01
|
E-13
E-14
E-15
|
E-16
Symptom
The rotor stops with
deceleration.
Rotor stops without
deceleration to standstill.
Instrument cannot be
operated.
Rotor stops with
deceleration to standstill.
Instrument cannot be
operated.
Rotor stops with
deceleration to standstill.
Instrument cannot be
operated.
Possible causes and corrective measures
Imbalance switch releases.
1. Open the instrument by pressing the "lid open"
2. Check whether the rotor is properly loaded.
3. Check whether a broken tube or a damage to the rotor
release the imbalance switch.
If you cannot locate and solve the problem yourself, contact a
service representative.
Internal program error.
Switch the instrument off and on again.
If the error persists, contact service representative.
Overtemperature in the centrifuge chamber.
Switch the centrifuge off and turn it on again after approx. one
minute.
If the error persists, contact service representative.
Temperature measurement error.
Switch the instrument off and on again.
If the error persists, contact service representative.
Troubleshooting
in case.
51