Preparations for operating the device
8.6

Tubing

Danger of injury from defective tubing.
The bath fluid tubing is a potential source of danger at high working
temperatures. Large volumes of hot bath fluids can be pumped out of a
damaged tubing in a short period of time.
Possible consequences:
Danger from unsealed pump connections.
Danger from the incorrect use of tubing.
We recommend that you select suitable tubing on the JULABO
homepage.
Pos: 2 8 /Technis che D okumentation/Flash CORI O/ CORIO_ CD_GESAMT /Inbetriebnahme - Netzs pannung, a usführliche Warnung @ 0\mod_ 14092 1845 0561_ 18.docx @ 6 763 @ 1 @ 1
www.calcert.com
15
10
20
5
25
0
30
Skin burns
Breathing problems due to the hot atmosphere
If the pump connections are not sealed, bath fluid may be pumped
out without any control.
Set the lever on the pump to internal circulation.
Unused pump connections must always be sealed with sealing
screws.
The tubing must be suitable for the pressure and temperature range
which results from operation and for the bath fluid (for example
silicon oil must not be used with silicon tubings).
Secure the tubing connections to prevent them sliding off. Use tube
clamps.
Do not kink the tubing. This will reduce throughput and may cause
the maximum pressure in the system to be exceeded (glass
reactor). The tube length should therefore be kept at a reasonable
level.
Prepare a maintenance plan.
Check tubing at regular intervals, at least once per year, for signs of
material fatigue (for example cracking)
The tubing must be replaced at regular intervals if they are in
constant use.