Errors
Error Types
-420
Query UNTERMINATED
-430
Query DEADLOCKED
-440
Query UNTERMINATED
after indefinite response
+2000
Offset calibration on channel 1 failed
+2002
Gain calibration on channel 1 failed
+2004
Interpolator calibration failed
+2005
Oscillator calibration failed
+2007
Measurement hardware calibration failed
+2008
Measurement interpolator calibration failed
+2009
No valid external timebase
+2010
External timebase failed during
measurement
+2011
Recall setup failed; hardware failure
Recall setup failed; empty register
+2012
Save setup failed
+2013
EEPROM failed
5-10
Indicates that a condition causing an UNTERMINATED Query error
occurred. For example, the Counter was addressed to talk and an incomplete
program message was received.
Indicates that a condition causing a DEADLOCKED Query error occurred.
For example, both input buffer and output buffer are full and the Counter
cannot continue.
Indicates that a query was received in the same program message after a
query requesting an indefinite response ( for example, *IDN? or *OPT?) was
executed.
:DIAGnostic:CALibration:INP:OFFS:AUTO ONCE failed.
:DIAGnostic:CALibration:INP:GAIN:AUTO ONCE failed.
:DIAGnostic:CALibration:INTerpolator:AUTO ONCE, *CAL?, or
:CALibration[:ALL]? failed.
:DIAGnostic:CALibration:ROSCillator:AUTO ONCE failed.
A measurement calibration failed on the last measurement.
Interpolator calibration failed on the last measurement; therefore, no valid
measurement was taken.
You have selected external reference and there is no external reference
applied to the rear-panel Ref In connector, or the external signal is not an
allowed frequency.
You have selected the auto reference mode and the Counter detected that
the external reference became invalid during the measurement. Therefore,
the current result is not valid, and the Counter switches to using the internal
reference.
*RCL failed.
*RCL failed because the register specified for recall is empty.
*SAV failed, or *RCL couldn't save to register 0.
A request to update a setting which is stored in the EEPROM
(:SYST:COMM:SER:TRAN:BAUD, :SYST:COMM:SER:TRAN:PAR,
:SYST:COMM:SER:TRAN:PACE, :SYST:COMM:SER:CONT:DTR,
:DISP[:WIND]:TEXT:RAD, :CAL[:DATA], or :DIAG:CAL: ...) resulted in a
hardware failure.