HP 53181A Programming Manual page 179

Frequency counter
Hide thumbs Also See for 53181A:
Table of Contents

Advertisement

Commands Reference
:CALCulate3 Subsystem
Related
Stats
Front-Panel
keys
:CALCulate3:AVERage:COUNt:CURRent?
Queries the current count (that is, the number of data values collected for statistical
computation).
• Numeric data transferred as ASCII bytes in <NR1> format.
Query Response
• Range is 0 to 1,000,000.
• If :CALC3:AVER[:STATe] is OFF, error -221 is generated.
• If the current measurement is Voltage Peaks, 0 is returned and error -221 is
generated.
• Query only.
Comments
• No statistics results exist until the :CALC3:AVER:COUN:CURR? is equal to
the specified :CALC3:AVER:COUN.
Related
Stats
Front-Panel
keys
:CALCulate3:AVERage[:STATe] <Boolean>
Sets or queries the statistics post-processing enable .
• Single ASCII-encoded byte, 0 or 1.
Query Response
• A value of 0 indicates OFF; a value of 1 indicates ON.
• *RST: OFF
Comments
• When this enable is ON, and :TRIG:COUN:AUTO is ON, and
[:SENS]:FUNC[:ON] is not a Voltage Peak function, then :INIT[:IMM]
initiates a complete block of measurements. See :TRIG:COUN:AUTO in this
chapter for specifics.
• When this enable is OFF, :INIT[:IMM] always initiates a single measurement.
4-21