Commands Reference
:DIAGnostic Subsystem
:DIAGnostic:CALibration:INPut[1]:OFFSet:AUTO ONCE | OFF
Calibrates the channel 1 input trigger OFFSet when the ONCE parameter is used.
Before sending this command, BE SURE to disconnect any input signal from
channel 1.
Query Response
A sequence of ASCII-encoded bytes: OFF
• The calibration values are stored in non-volatile memory, and are unaffected by
Comments
power-on, save/recall, and *RST.
• Use :DIAG:CAL:STAT? to check for successful calibration.
• After calibration is completed, the state of this command's parameter is OFF.
Related
Scale & Offset/POWER (Calibration Menu)
Front-Panel
Keys
:DIAGnostic:CALibration:INTerpolator:AUTO ONCE | OFF |ON
Calibrates the interpolator circuit in the Counter when the ONCE parameter is used.
AUTO ON enables automatic interpolator calibration on every measurement.
AUTO OFF disables automatic interpolator calibration.
Query Response
A sequence of ASCII-encoded bytes: OFF or ON
• *RST: ON
Comments
• Use :DIAG:CAL:STAT? to check for successful calibration.
• After ONCE calibration is completed, the state of this command's parameter is
OFF.
• When :DIAG:CAL:INT is set to OFF, the Counter reports the positive status
condition (questionable Time and Frequency) to bits 2 and 5 of the
Questionable Status Register. When :DIAG:CAL:INT is set to ON, the Counter
reports the negative status condition (NOT questionable Time and Frequency)
to bits 2 and 5 of the Questionable Status Register.
• This enable is unaffected by save/recall.
:DIAGnostic:CALibration:ROSCillator:AUTO ONCE | OFF
Calibrates the reference oscillator when ONCE parameter is used.
4-33