Keysight Technologies N5250C PNA Series Manual page 47

Microwave network analyzer system
Table of Contents

Advertisement

PNA Series Microwave Network Analyzer System
N5250A
Figure 3-12
New Design Combiner Assembly and mm-Wave Module: View 3
(4 places) 2
4. Remove the four waveguide screws (item 1).
5. Remove the four screws (item 2).
6. Reverse the process to reassemble.
Do NOT switch modules. Combiner must be reassembled to the mm-wave
module in which it was originally installed.
7. Perform an Operator's Check after reassembly to test for power holes, etc.
Installation and Service Guide N5250-90001
Disconnecting the Combiner Assembly From the mm-Wave Module
1 (4 places)
System Installation
N5230_013_318
3-21