LOGIQ 7/LOGIQ 7 Pro Quick Guide

Declaration of Immunity

This system is suitable for use in the following environment. The user must assure that the system is used according to the specified guidance and only in the
electromagnetic environment listed.
Immunity Type
IEC 61000-4-2
Static discharge
(ESD)
IEC 61000-4-4
Electrical fast
transient/burst
IEC 61000-4-5 Surge Immunity
IEC 61000-4-11
Voltage dips, short interruptions and
voltage variations on mains supply
IEC 61000-4-8
Power frequency (50/60 Hz) magnetic
field
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
Radiated RF
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. If noise generated from other electronic
equipment is near the probe's center frequency, noise may appear on the image. Good power line isolation is required.
Direction 5307393-100 Rev. 1
Table 1-7: Declaration of Immunity
Test Level
± 6 kV contact
± 8 kV air
± 1 kV for mains
± 0.5 kV for ECG cable
± 1 kV differential
± 2 kV common
< 50
(> 95% dip) for 0.5 cycle;
T
400
(60 0ip) for 5 cycles;
T
700
(30 0ip) for 25 cycles;
T
< 50
(>95% dip) for 5 sec
T
3 A/m
3 V
RMS
150 kHz - 80 MHz
3 V/m
80 MHz - 2.5 GHz
Compliance
± 6 kV contact
± 8 kV air
± 1 kV for mains
± 0.5 kV for SIP/SOP
± 1 kV differential
± 2 kV common
< 50
(> 95% dip) for 0.5 cycle;
T
400
(60 0ip) for 5 cycles;
T
700
(30 0ip) for 25 cycles;
T
< 50
(>95% dip) for 5 sec
T
3 A/m
3 V
RMS
150 kHz - 80 MHz
3 V/m
80 MHz - 2.5 GHz
EMC Environment and Guidance
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial
and/or hospital environment. If the user requires continued
operation during power mains interruptions, it is
recommended that the system be powered from an
uninterruptable power source (UPS) or a battery.
NOTE: UT is the a.c. mains voltage prior to application of the
test level.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial and/
or hospital environment.
Separation distance to radio communication equipment must
be maintained according to the Table 1-5 on page 46.
Interference may occur in the vicinity of equipment marked
with the symbol
50