Wiring Plan, Single Chamber Housing - Siemens SITRANS LG240 Operating Instructions Manual

Guided wave radar
Hide thumbs Also See for SITRANS LG240:
Table of Contents

Advertisement

PBD-51041037
Fig. 10: Connection steps 5 and 6
1
Single chamber housing
2
Double chamber housing
6. Insert the wire ends into the terminals according to the wiring plan
Note:
Solid cores as well as flexible cores with wire end sleeves are insert-
ed directly into the terminal openings. In case of flexible cores without
end sleeves, press the terminal from above with a small screwdriver,
the terminal opening is then free. When the screwdriver is released,
the terminal closes again.
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
external ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing lid back on
The electrical connection is finished.
5.3

Wiring plan, single chamber housing

The following illustration applies to the non-Ex, Ex-ia and Ex-d ver-
sion.
SITRANS LG240 - Operating Instructions
1
2
21

Advertisement

Table of Contents
loading

Table of Contents