Work-Up Of Holes - Siemens MAMMOMAT 3000 Modification Instructions

Collimator holder
Hide thumbs Also See for MAMMOMAT 3000:
Table of Contents

Advertisement

2-4

Work-up of holes

1
Remove collimator protective cover by unscrewing four Allen screws.
Figure 4 Removal of collimator protective cover
2
Disconnect cables X866, X867 and X890 from the collimator and cut off the cable
ties. Notice the connector and cable-tie positions to ensure the same positioning after
work-up.
3
Unscrew the four Allen screws holding the collimator with holder.
Figure 5 Removal of collimator holder
4
Fold down the collimator with holder and secure with a cable tie to prevent it from
hanging in the cables.
5
Cover the collimator with a plastic bag. Protect other parts of the Mammomat with a
plastic cover.
Siemens-Elema AB
Solna, Sweden
Modification of collimator holder
SRX-MK
RX B7-230.092.01.01.02
REV.01
10.95
MAMMOMAT 3000