Installing the paneling
Reattach the protective plate behind the keyboard in the logbook compartment.
Take care that the cables are not squeezed.
Refit the paneling of the logbook compartment. Make sure the ground wire connections
are correct when doing this.
Reinstall the rear paneling of the monitor trolley. Make sure the ground wire connections
are correct when doing this.

Final work

After all work has been completed and all covers have been attached perform the protec-
tive conductor test according to ARTD-002.731.17... The protective conductor resis-
tance must not exceed 0.2 Ohm.
Add these instructions to the system manual of the product.
File the Addendum/Supplement to the Operating Instructions included in the scope of
supply into the existing operating instructions.
Customer information
Demonstrate the function of the key switch.
Hand over the keys.
Hand over the operating instructions.