When the I-CAL function is activated, the internal adjustment is
I-CAL
started only by pressing the CAL button. Before pressing CAL
ensure that there are no objects on the weighing plate.
At the models with internal adjustment weight the adjustment with
E-CAL
external weight is not recommended.
Procedure see chapter 7.1.

8.2.1 Adjustment models PLJ-M

In weighing mode press the MENU button and keep it pressed until
Menu settings
the acoustic signal gets mute. The first menu item „units" is displayed.
Press MENU button repeatedly until „calib" appears and confirm using
PRINT button.
Use the MENU key to choose between the following settings:
AUT-CAL
Take over selection using the PRINT button.
The balance returns to menu.
To finish the menu press the MENU button and keep it pressed until
the acoustic signal gets mute. The balance returns automatically into
weighing mode.
ALJ/ALS/PLJ/PLS-BA-e-1943
Automatic adjustment with internal weight.
Factory setting for models with type approval.
37