6
Maintenance
User maintenance of the Polaris Vicra System is limited to the following procedures:
1. Cleaning the Position Sensor
2. Care of cables and connectors
The Position Sensor is the only system unit that requires special consideration. The power adapter
and Host USB Converter can be cleaned by following standard procedures.
Maintenance Warnings
Before doing any maintenance on the Polaris Vicra System, read the following warnings:
1. All user maintenance must be done by appropriately trained personnel. Individual
components of the Polaris Vicra System contain no user-serviceable parts. Maintenance by
Warning!
untrained personnel may present an electric shock hazard.
2. Do not use the Position Sensor without inspecting it for cleanliness and damage both
before and during a procedure. Reliance on data provided by an unclean or damaged
Position Sensor may lead to inaccurate conclusions. If your application involves personal
safety, inaccurate conclusions increase the possibility of personal injury.
3. Do not immerse any part of the Polaris Vicra System or allow fluid to enter the equipment.
If fluids enter any part of the system they may damage it and present a risk of personal
injury.
4. Do not sterilize the Polaris Vicra Position Sensor as this may cause irreversible damage to
its components. Reliance on data provided by a damaged Position Sensor may lead to
inaccurate conclusions. If your application involves personal safety, these inaccurate
conclusions increase the possibility of personal injury.
6.1

Position Sensor

Regularly inspect the Position Sensor for cleanliness. The Position Sensor and particularly the
illuminator filters and lenses, should be cleaned only when necessary. The frequency of cleaning
must be determined by the user. This may include "in-use" cleaning.
Caution!
Use only 70% isopropanol and a lens cleaning solution formulated for multi-coated lenses (for example, AR66) to
clean the Position Sensor. Other fluids may cause damage to the illuminator filters. Do not use any paper
products for cleaning. Paper products may cause scratches on the illuminator filters.
To clean the Position Sensor, follow the procedure detailed below:
1. Remove dust from each illuminator filter and lens, using a photographic lens duster (brush).
Gently wipe the surface in one direction only, by pulling the brush across the surface.
2. Gently wipe the illuminator filters and lenses with disinfectant wipes containing 70%
isopropanol. Continue cleaning the remainder of the Position Sensor, taking care not to wipe
52