Using the Polaris Vicra System
4. Click
5. Select a start delay of about 5 seconds and a duration of about 20 seconds.
To Pivot the Tool
1. Place the tool tip in the divot.
2. Ensure that the tool is within the characterized measurement volume, and will remain within the
volume throughout the pivoting procedure.
3. Click Start Collection in the Pivot tool dialog.
4. Pivot the tool in a cone shape, at an angle of 30º to 60º from the vertical.
a) Keep the tool tip stationary, and ensure that there is a line of sight between the markers on
the tool and the Position Sensor throughout the pivoting procedure.
b) Pivot the tool slowly until the specified pivot duration time has elapsed.
46
Figure 4-7 Selecting a Tool to Pivot
to open the Pivot dialog.
Polaris Vicra User Guide - Revision 2