Characteristic Performance

Characteristic performance for RF and microwave ECal modules are provided in the
following tables, which describe non-warranted performance that most units exhibit.

8509x Series

Characteristic Performance for 85091C (7 mm)
Parameter
Directivity (dB)
Source Match (dB)
Reflection Tracking (±dB)
Transmission Tracking
Load Match (dB)
Characteristic Performance for 85092C (Type-N 50Ω)
Parameter
Directivity (dB)
Source Match (dB)
Reflection Tracking (±dB)
Transmission Tracking
Load Match (dB)
Characteristic Performance for 85093C (3.5 mm)
Parameter
Directivity (dB)
Source Match (dB)
Reflection Tracking (±dB)
Transmission Tracking
Load Match (dB)
Chapter 5
300 kHz to 1 GHz
52.0
45.0
0.04
±
0.06
(
dB)
47.0
300 kHz to 1 GHz
52.0
45.0
0.04
±
0.04
(
dB)
47.0
300 kHz to 1 GHz
52.0
44.0
0.03
±
0.04
(
dB)
47.0
Specifications and Characteristics
Frequency Range
1 to 3 GHz
3 to 6 GHz
56.0
55.0
44.0
41.0
0.04
0.07
0.06
0.13
47.0
46.0
Frequency Range
1 to 3 GHz
3 to 6 GHz
54.0
52.0
44.0
41.0
0.04
0.06
0.04
0.07
47.0
44.0
Frequency Range
1 to 3 GHz
3 to 6 GHz
52.0
50.5
44.0
39.0
0.03
0.05
0.05
0.07
47.0
44.0
Characteristic Performance
6 to 9 GHz
45.0
34.0
0.1
0.23
39.0
6 to 9 GHz
47.0
36.0
0.07
0.14
39.0
6 to 9 GHz
47.0
34.0
0.07
0.12
40.0
5-5