twelve 1-port standards and three 2-port (thru) standards are measured, requiring a total
of eighteen connections. The same calibration using a 4-port ECal module (N4431B,
N4432A and N4433A) requires only four connections.
Ignoring connection related errors, the following table compares the accuracy of ECal
versus mechanical calibration in various calibration options.
Table 1-1
Mechanical Calibration Type
1-Port, calibration constant (not
data-based) SOL
1-Port, short, open, sliding load
1-Port, data-based, expanded
2-Port TRL
≥2-Port, calibration constant
a
SOL, flush thru
≥2-Port, calibration constant
SOL, unknown thru
≥2-Port, data-based, expanded,
a
flush thru
≥2-Port, data-based, expanded,
unknown thru
++ better; + slightly better; = same
a. Flush thru should NOT be used when the test port connectors are not the same type as the
calibration standards. SMA and 3.5 mm connectors are mateable but are NOT the same
type.
Chapter 1
ECal Thru
Measurement
Type
NA
Reflection
NA
Reflection
NA
Reflection
All types
Reflection
Transmission
characterized
Reflection
Transmission
internal
Reflection
unknown
Transmission
external
Reflection
a
Transmission
flush
characterized
Reflection
Transmission
internal
Reflection
unknown
Transmission
characterized
Reflection
Transmission
internal
Reflection
unknown
Transmission
external
Reflection
Transmission
a
flush
characterized
Reflection
Transmission
internal
Reflection
unknown
Transmission
General Information
Manual Overview
ECal
Mechanical
Calibration
++
+
+
++
++
++
++
++
++
++
++
++
=
=
+
++
+
+
+
+
+
++
+
+
1-3