Electrical Connections - Siemens LDS 6 Operating Instructions Manual

In-situ laser gas analyzer
Table of Contents

Advertisement

Installation Guidelines

3.2 Electrical connections

3.2
Electrical connections
Transmitter
Install the cables, hybrid and loop cable, at the transmitter sensor. Cables are fixed in the
sensor with gaskets. Always use lubricant on the gaskets before installing, otherwise these
will be very difficult to remove later. Multi-mode fiber (SMA connector) and 24 V are just
connected straight through to the receiver.
Note
Always protect the single-mode fiber (E2000 connector) from dust. Do not remove the plastic
vial until the cable is firmly secured in the sensor.
Figure 3-3
Receiver
Install the loop cable in the receiver sensor. Cables are fixed with gaskets. Always use
lubricant on the gaskets before installing, otherwise they will be very difficult to remove later.
Connect the multi-mode fiber to the sensor-board as shown below.
Figure 3-4
16
Connections in the transmitter
Connections in the receiver
Sensor CD 6C Operating Instructions
Operating Instructions, 12/2007, A5E00912721-02