Installation Guidelines

3.4 ATEX considerations

3.4
ATEX considerations
A hybrid cable is used to connect the transmitter sensor to the barrier.
Connecting Transmitter Sensor to Barrier
1. Cut off the pre-mounted phoenix contact at the hybrid cable.
2. Insert the hybrid cables (one for each channel) in the feed throughs at the top of the
barrier box. Secure each cable with a gasket.
3. Connect each hybrid cable to the Surge Protection Devices SD32X (yellow). - (brown
cable) shall be connected to inlet number 4 and + (green cable) to inlet number 5.
4. Connect the hybrid cable to the transmitter sensor.
Connecting Receiver to Transmitter Sensor
For ATEX versions of the LDS 6 the receiver sensor are pre-mounted with special ATEX
electronics and surge protection. The green phoenix contact mounted on the sensor card is
already connected with a red wire for +24 V and a black wire for GND. Two additional wires
should be connected to the phoenix connector. These wires come from the loop cable.
1. Connect the green wire next to the red one (for 24 V) on the phoenix contact.
2. Connect the brown wire next to the black one (for GND) on the phoenix contact.
Connect the sensor housings, on both the transmitter and the receiver side, to earth using
the pre-mounted 2 meters earth cable (yellow-green).
See also
The CD 6C Ex sensor (Page 29)
24
Sensor CD 6C Operating Instructions
Operating Instructions, 12/2007, A5E00912721-02