SPECIFICATIONS

RESPIRATION AND RR

Table 10 indicates the Specifications associated with Respiration and RR Parameter.
Patient Populations
Method
Input Connector
Sensing Lead
RR Resolution
Measurement Range
Measurement Accuracy
Measurement Sensitivity
Report Interval
Bandwidth
Impedance Measuring Current
PARAMETER ALARM LIMITS & SETTINGS
Parameter Units
Alarms On
Default
HR upper
bpm
Yes
HR lower
bpm
Yes
RR upper
rpm
Yes
RR lower
rpm
Yes
Note:
Lower Alarm Limit cannot be set above the associated Upper Alarm Limit.
Note:
Upper Alarm Limit cannot be set lower than the associated Lower Alarm Limit.
21-22-0335 Rev D
Adult, adolescent, child and infants
Impedance Pneumography
Same as ECG
II
1 rpm (breaths per minute)
0 to 120 rpm
±3 rpm
0.25 ohms (minimum)
1 second
0.17 to 3.3 Hz (-3dB)
40 μA @ 28 kHz square wave across Lead II
Table 10: Respiration & RR Specifications
Adult
Alarms
Limit
Limit
Default
Range
Default
16 – 300
120
50
15 - 299
20
6 - 150
5
5 - 149
Table 11: HR & RR Parameter Alarm Limit Ranges
Pediatric
On
Limit
Limit
Default
Range
16 – 300
Yes
120
Yes
50
15 - 299
Yes
45
6 - 150
Yes
10
5 - 149
740
SELECT
Neonatal
Alarms
On
Limit
Limit
Default
Default
Range
16 – 300
Yes
180
Yes
100
15 - 299
NA
NA
NA
NA
NA
NA
Page 38 of 40