Electrode Placement; Standard Placement (Aha); Standard Placement (Iec); Figure 5: Standard Aha Electrode Placement - ZOE 740 SELECT User Manual

Multi-parameter monitor
Table of Contents

Advertisement

ECG MONITORING

ELECTRODE PLACEMENT

STANDARD PLACEMENT (AHA)

For ECG cable lead sets with AHA (USA) lead designations, position the electrodes as
shown in Figure 5 below:
RA (White)
1. Position the right arm (RA) electrode on the right mid-clavicular line, directly below the
clavicle.
2. Position the left arm (LA) electrode on the left mid-clavicular line, directly below the
clavicle.
3. Position the left leg (LL) electrode on the left mid-clavicular line, 6
space.

STANDARD PLACEMENT (IEC)

For ECG cable lead sets with IEC (Europe) lead designations, position the electrodes as
shown in Figure 6 below:
R (Red)
1. Position the right arm (R) electrode on the right mid-clavicular line, directly below the
clavicle.
2. Position the left arm (L) electrode on the left mid-clavicular line, directly below the clavicle.
3. Position the left leg (F) electrode on the left mid-clavicular line, 6
space.
21-22-0335 Rev D

Figure 5: Standard AHA Electrode Placement

Figure 6: Standard IEC Electrode Placement

740
SELECT
LA (Black)
LL (Red)
th
th
and 7
intercostal
L (Yellow)
F (Green)
th
th
and 7
intercostal
Page 16 of 40