CLEANING

8. CLEANING
Caution:
Always disconnect the monitor from AC mains before cleaning.
Do not use harsh chemicals for cleaning – in particular, do not use disinfectants that
Caution:
contain phenol as they can spot plastics. Do not steam autoclave, gas sterilize, irradiate,
subject to intense vacuum, or immerse in water or cleaning solution. Be careful to avoid
getting cleaning liquids into connectors or the unit. If this occurs, allow the unit to dry in warm
air for 2 hours, then check to make sure all monitoring functions are still working properly.
Caution:
Accessories that fall on the floor should be inspected for contamination and proper
functionality. If contamination is observed, then this cleaning procedure should be followed.
Caution:
The user has the responsibility to validate any deviations from the recommended
method of cleaning and disinfection.
Table 8 provides cleaning instructions for ECG Accessories used with the monitor, which should
be cleaned monthly or as warranted. Accessories should be cleaned before application to a
patient. Before cleaning, refer to the cautions listed above.
Part
Materials
Enzymatic detergent such as ENZOL
Distilled water
Disinfectant solution (such as CIDEX OPA
bleach (5.25% sodium hypochlorite) in distilled water)
Soft cloths and/or soft-bristled brushes
Protective gloves and eyewear
Procedure
1.
2.
ECG
3.
Cables
4.
5.
6.
7.
8.
9.
10. Wipe away excess solution and clean product again with cloth dampened in
11. Allow 2 hours for drying.
1
ENZOL is a Registered Trademark of Synclaire Brands, Inc.,
2
CIDEZYME and CIDEX OPA are a Registered Trademarks of Advanced Sterilization Products (ASP)
21-22-0335 Rev D
Recommended cleaning method
Disconnect the unit from the wall outlet.
Put on gloves and protective eyewear.
Prepare the enzymatic detergent according to the manufacturer's instructions,
and also the disinfectant solution, in separate containers.
Apply detergent to product using a soft cloth - If material is dried on, allow to sit
for 1 minute.
Wipe smooth surfaces with the cloth.
Use a soft-bristle brush on visibly soiled areas and irregular surfaces.
Remove detergent from product using cloth dampened in distilled water.
Repeat as necessary.
Apply disinfectant solution on affected area using a soft cloth - Allow product to
sit for 5 minutes.
distilled water.

Table 8: ECG Accessory Cleaning Instructions

® 1
(US) or CIDEZYME
® 2
, or a 10% solution of household
740
SELECT
® 2
(outside the US)
Page 36 of 40