Table 12: Ecg, Hr & Rr Parameter Setting Ranges - ZOE 740 SELECT User Manual

Multi-parameter monitor
Table of Contents

Advertisement

SPECIFICATIONS
Table 12 lists the ECG, HR & RR Parameter settings and Ranges.
Parameter
ECG Lead
(ECG Lead) Size
HR Source
Pulse Tone Source
ECG Filter
Channel 1-3 Waveform
ECG Waveform Size
Sweep Speed
Show Beat Detect Spikes
Notch Filter
Table 12: ECG, HR & RR Parameter Setting Ranges
21-22-0335 Rev D
Defaults
Adult
Pediatric
II
II
10 mm/mV
10 mm/mV
Auto
Auto
Off
Off
Monitor
Monitor
(0.67 – 40 Hz)
(0.67 – 40 Hz)
II
II
10 mm/mV
10 mm/mV
25 mm/sec
25 mm/sec
Yes
Yes
60 Hz
60 Hz
Neonatal
II
I, II, III
10 mm/mV
2.5, 5, 10, 15 mm/mV
Auto
Auto, ECG, SpO
Off
ECG, SpO
Monitor (0.67 – 40 Hz),
Monitor
(0.67 – 40 Hz)
Diagnostic (0.05 – 40 Hz)
II
I, II, III
10 mm/mV
2.5, 5, 10, 15 mm/mv
25 mm/sec
6.25, 12.5, 25 mm/sec
Yes
No, Yes
60 Hz
60 Hz, 50 Hz, Off
740
SELECT
Range
, NIBP
2
, Off
2
Page 39 of 40