ECG MONITORING
ANALYZE PACERS
Touch the On adjacent to the Analyze Pacers selection to Enable.
OR
Touch the Off adjacent to the Analyze Pacers selection to Disable.
When Analyze Pacers is set to On, the monitor will display white pace detect marks at the top
of the ECG waveform to indicate where pacer pulses were detected.
When Analyze Pacers is set to Off, the the monitor will display a message advising you that
the pacer detect function has been disabled.
Warning:
PACEMAKER PATIENTS. When Analyze Pacers is disabled, the Heart Rate may
continue to count the pacemaker rate during occurrences of cardiac arrest. Keep pacemaker
patients under close surveillance and do not rely entirely upon Heart Rate alarms.
PULSE TONE
The user-selectable Pulse Tone feature allows the clinician to select the desired to sound a
tone for each pulse detected.
User Selectable settings: Off (Default) & On
ECG LEAD
The user-selectable ECG Lead feature allows the clinician to select the desired ECG Lead
waveform on the display.
User Selectable settings: I, II (Default) & III
SIZE
The user-selectable Size feature allows the clinician to select the desired size of the ECG
waveform on the display.
User Selectable settings: 2.5, 5, 10 (Default), 15 & 20 mm/mV