INTENDED USE AND INTRODUCTION

2. INTENDED USE

INTENDED USE
Full Indications for Use and Contraindications may be found in the
Manual, Zoe Medical PN 21-22-0316.
The
740
monitor facilitates the monitoring of:
SELECT
ECG and heart rate (HR) with the ECG option for adult and pediatric patients;
Impedance pneumography to detect respiration rate (RR) or absence of respiratory effort
with the ECG option for adult, adolescent, child and infant patients.
The
740
monitor is a prescription device intended to be used by healthcare professionals
SELECT
in all areas of a healthcare facility.

OVERVIEW OF HEART RATE MONITORING

ECG Monitoring works through the sensing the electrical signals generated by the electrical
activity of the heart as it beats. These signals are acquired from chest electrodes, and the
monitor amplifies the signals so they can be displayed on the screen. The patient's heart
SELECT
rate (HR) is calculated and continuously updated based on running average of the R to R
intervals between each QRS complex.
The ECG and HR monitoring capabilities of the
3-wire cable monitoring capabilities
Calculating the average heart rate (HR) in beats per minute
Detecting asystole and ventricular fibrillation
Pacer pulse detection
Generating an audible pulse tone for each detected beat