Inspection and maintenance
2.7

Double-slot diaphragm

Connect the power plug, POLYMOBIL ON.
PMF
Light localizer lamp
Check if the light localizer lamp works, and replace it if necessary.
See "Changing the light localizer lamp" in the service instructions.
SIM
Rotatability
Check if the double-slot diaphragm can be turned easily at ± 45° (Fig.4).
PMF
Format setting
By turning the collimator adjustment grips, check for easy movement and evenness of
the format setting (pairs of plates).
POLYMOBIL OFF and disconnect the power plug.
PMF
0° position
Move the support arm to the top position.
Use the spirit level on the double-slot diaphragm to check the 0° position. If necessary,
readjust the scale on both sides.