2 - 8
PMP
Maintenance of the joints and chains
Spray the stand joints with multi-purpose WD-40 spray.
Lubricate the chains (5/Fig.8) using Optimol Longtime PD2 grease.
Attach cover (3/Fig.7) to the stand.
2.9

Radiation

Connect the power plug, POLYMOBIL ON.
Close the double-slot diaphragm, and specify 60 kV, 10 mAs.
SIE
Radiation indicator
SIE
Acoustic signal
Trigger exposure.
The radiation indicator (1/Fig.9) must be lit during exposure, and an acoustic signal will
sound at the same time.
POLYMOBIL OFF and disconnect the power plug.
Polymobil 10
Inspection and maintenance
Fig. 9
RXR8-120.101.01
Rev. 05
07.05
Page 8 of 14
CS PS 24
1
Siemens AG
Medical Solutions