Siemens Polymobil 10 Maintenance Manual page 14

Table of Contents

Advertisement

1 - 10
WARNING
The system must be switched on during the measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
Enter the highest value into the leakage current report.
This value must not exceed the permissible leakage current values of 2.5 mA according
to DIN VDE 0751-1/2001-10, Table F.1, row "System leakage current for devices
according to remarks 1 and 3".
Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, correct the measured value to the value corresponding to a measure-
ment at the nominal value of the line voltage. This is also to be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
In the case of repeat measurements, the measured value must also be evaluated.
NOTE
The evaluation is not necessary in the case of a new determination.
File the leakage current report in the "Certificates" register in the system binder.
NOTE
NOTE
Polymobil 10
Electrical voltage!
Non-compliance can lead to severe injury and even death.
No housing parts of the system may be touched during direct
measurement of the system leakage current (measurement setup
according to Fig. 4).
Access to the system by third parties must be prevented.
Evaluate the results by comparing the first measured value to the
corresponding values documented during preceding mainte-
nance procedures or safety checks.
A sudden or unexpected increase of the measured values may
indicate that a fault has occurred in the primary circuit of the
power supply (damaged insulation, damage caused by humidity,
defective interference suppressor, etc.) - even if the limit value
of 2.5 mA is not exceeded.
The leakage current report is filed in the "Certificates" register
in the system binder.
A new report must be created if the leakage current measure-
ments are not documented.
These instructions contain a new report called 'Leakage current/
report'. Separate this report, fill it out, and file it in the
"Certificates" register in the system manual. If values are newly
determined, they are to be recorded as first measured values.
The evaluation is omitted in this case.
RXR8-120.101.01
Rev. 05
General information
Page 10 of 14
07.05
CS PS 24