Checking coincidence of light field and radiation field
Connect the power plug, POLYMOBIL ON.
Load a 24 cm x 30 cm or 10" x 12" cassette with film and place it on a table or similar
support.
Use a tape measure to set a vertical SID of 100 cm or 40" to the top edge of the cassette.
Use the control knobs to set a format of 18 cm x 24 cm or 8" x 10".
Switch the light localizer on and align the cassette.
Place radio-opaque marks (e.g. washers, coins) on the cassette as shown in the
diagram.
Use a washer as side marker.
Release exposure (60 kV, 10mAs), and process the film.
Make a note on the processed film of the following data using a waterproof felt tip pen:
- Adjusted SID
- Film size
- Radiation field size