MAGSTIM BiStim2 Operating Manual page 9

Hide thumbs Also See for BiStim2:
Table of Contents

Advertisement

3.1.3
The power indicator flashes when the Magstim Bistim² is in standby, and illuminates when the
Magstim Bistim² is on. When power is initially applied to the Magstim Bistim², the power indicator
should flash indicating the unit is in standby state.
3.1.4
The output display consists of 3 lines. The first and second lines display the power output as a
percentage of maximum of the master and slave units respectively. The third line displays the inter-
pulse spacing. The default value displayed after initially switching on the master and slave units are
30% for both power outputs and 10ms for the inter-pulse spacing.
If an error code needs to be displayed, the power output is replaced by error code(s). These are
detailed in Appendix A – System Error Codes.
3.1.5
This green LED will illuminate if there is a coil detected as being connected correctly to the Magstim
Bistim².
3.1.6
These LED's represent the surface temperature of the coil. The LED's illuminate sequentially as the
coil temperature rises, beginning with the bottom green LED and ending with the amber LED. When
the amber LED is lit, it means that the coil is getting close to its maximum operating temperature. As
soon as it reaches its maximum operating temperature, the replace coil indicator will illuminate and
the Magstim Bistim² will disarm itself.
Note: If the temperature of the coil is below 5
will remain in its inactive default condition until it detects that the coil temperature has risen above
o
5
C.
3.1.7
This red LED will illuminate if any of the following conditions are present:
1.
2.
3.
4.
If this LED is illuminated the Magstim Bistim² will not arm until the reason for the indicator being lit is
resolved.
© The Magstim Company Limited
Power Indicator* (3)
Output Display* (4)
Coil Indicator* (5)
Coil Temperature Indicators* (6)
Replace Coil Indicator* (7)
The coil connected is not compatible with the Magstim Bistim².
The surface temperature of the connected coil has risen above its set limit.
The connected coil is faulty.
There is no coil connected.
o
C, the system will not operate. The Magstim Bistim²
6
Magstim BiStim²
Operating Manual
MOP02-EN-02