3.1.8
This LED is illuminated when the Magstim Bistim² has been put into its safe disarmed state.
3.1.9
When the Magstim Bistim² is armed, this LED will illuminated when the Magstim Bistim² is charged
and ready to be triggered.
3.1.10
The Magstim Bistim² can be armed pushing the green run button. The Magstim Bistim² will charge to
the power level indicated on the output display. The Magstim Bistim² will not arm if the replace coil
indicator is illuminated.
3.1.11
The output control allows the master power level, slave power level and inter-pulse spacing displayed
on the master unit UI to be adjusted. The master output power, slave output power and inter-pulse
spacing can be adjusted by rotating the output control clockwise to increases the level and rotating
anti-clockwise to decreases the level. To toggle between the master power, slave power and inter-
pulse spacing press the output control.
Each complete 360º rotation changes the master/slave power output level by approximately 20%.
3.1.12
The Magstim Bistim² can be triggered to deliver the stimulus by pressing the yellow trigger button
along with one of the trigger buttons located on the coil. The instrument will continuously charge and
discharge if the trigger buttons are pressed and held.
3.1.13
The Magstim Bistim² can be disarmed by pushing the red stop button. In this mode the disarmed
indicator will illuminate and the instrument will discharge internally.
Note: The Magstim Bistim² will disarm if the coil is disconnected, becomes faulty, the surface
temperature of the connected coil has risen above its set limit, the system stop button is activated, the
instrument detects an internal fault, or if the ready symbol has been illuminated for over 1 minute
without the instrument having been triggered.
3.1.14
This socket is used to connect the stimulating coil to the Magstim Bistim² system.
© The Magstim Company Limited
Disarmed Indicator* (8)
Ready Indicator* (10)
Run Button* (11)
Output Control* (12)
Trigger Button* (13)
System Stop Button* (14)
Coil Output Socket* (15)