The Magstim Bistim² is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment
should assure that it is used in such an environment.
Immunity Test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-4
Electrical fast
± 2 kV for power
Transient/ burst
supply lines
IEC 61000-4-11
± 1 kV for input/
output lines
Surge
± 1 kV differential
mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
0,5 cycle
on power supply
input lines.
40 % U
(60 % dip in U
IEC 61000-4-11
cycles
70 % U
(30 % dip in U
cycles
<5 % U
(>95 % dip in U
sec
Power Frequency
3 A/m 50Hz
(50/60Hz)
Magnetic Field
IEC 61000-4-8
80MHz – 2.5GHz
Radiated RF
Immunity
2Hz 80% amplitude
EN61000-4-3
modulation
0.15MHz – 80MHz
Conducted RF
Immunity
2Hz 80% amplitude
EN61000-4-6
modulation
NOTE
U
is the a.c. mains voltage prior to application of the test level
T
© The Magstim Company Limited
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC 60601
Compliance
Test Level
Level
Meets requirement
Meets requirement
Meets requirement
Meets requirement
T
) for
T
T
) for 5
T
T
) for 25
T
T
) for 5
T
Meets requirement
Meets requirement
Meets requirement
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the Magstim Bistim² requires
continued operation during power interruptions, it is
recommended that the Magstim Bistim² be powered from an
interruptible power supply.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment
Equipment should only be used in the vicinity of other equipment
compliant with EN60601-1-2.
Equipment should only be used in the vicinity of other equipment
compliant with EN60601-1-2.
27
Magstim BiStim²
Operating Manual
MOP02-EN-02