Starting The Pump - Varian Turbo-V 3KT Instruction Manual

Rack controller
Table of Contents

Advertisement

Press PUMP CURRENT to confirm. and the
display shows:
P U M P
L
I
R E S E T ?
Turbo-V 3K-T
where:
PUMP LIFE = is the elapsed operating time range
000 to 99999 hours.
RESET XXX = YES or NO.
The controller is factory-set to NO.
When PUMP LIFE is reset to 000, the CYCLE number
is also reset to 000.
F E
X X X X X h
X X X
NOTE

Starting the Pump

If the forepump and vent device are not operated by the
controller, close the vent valve and switch on the
forepump.
Before starting the pump the cooling mode must be
selected.
With the FRONT panel operation selected, the
REMOTE and RS 232/485 operations are inoperative;
conversely,
the
CURRENT push-buttons are always active, even when
the
operating
mode
RS 232/485.
S O F T
S T A R T
[
-
-
-
Where the sign minus (-) become a square ( ) when the
pump finish the ramp-up step. The active step is indicated
by a flashing square ( ).
As the ten steps are fully covered, the pump will reach the
Normal operation. If during the Soft Start mode the power
drawn by the pump exceed 200 W the speed of the pump
is decreased to maintain the maximum power allowable
(200 W).
88
TECHNICAL INFORMATION
NOTE
NOTE
CYCLE
NUMBER
and
selected
is
REMOTE
O N
-
-
-
-
-
-
]
Turbo-V 3K-T
87-900-988-01 (B)
PUMP
or

Advertisement

Table of Contents
loading

This manual is also suitable for:

869-9447869-9547

Table of Contents