Table of Contents Safety Information ..............................3 Alert Signals..............................3 Unpacking ................................4 Visible Loss or Damage..........................4 Concealed Loss or Damage ........................4 Performance Characteristics ..........................5 Temperature Ranges ............................5 Electrical Requirements..........................5 Installation ................................6 Selecting a Location ............................6 Leveling The Unit ............................6 Shelves ................................6 Electrical Connection ............................7 Operation ................................8 Initial Startup ..............................8 Temperature Control ............................8...
Safety Information Your satisfaction and safety are important to Thermo Scientific and a complete understanding of this unit is necessary to attain these objectives. As the ultimate user of this apparatus, it is your responsibility to understand its proper function and operational characteristics.
Unpacking Save all packing material if apparatus is received dam- aged. This merchandise was carefully packed and thor- oughly inspected before leaving our factory. Responsibility for its safe delivery was assumed by the carrier upon acceptance of the shipment; therefore, claims for loss or damage sustained in transit must be made upon the carrier by the recipient as follows: Visible Loss or Damage...
Installation S electing a Location Choose a location for the refrigerator that will provide at least three inches of clearance between the cabinet and any adjacent vertical surface at the sides and rear. Twelve inches of clearance is required above the grille. Appropriate electrical power must be available.
NSTALLATION Electrical Connection Caution Determine the total amount of current presently being Be sure that the power supply is the used by other apparatus connected to the circuit that will same voltage that is specified on the be used by this refrigerator. It is critical that this added refrigerator’s data plate.
3000 ft., have the thermostat adjusted by a techni- cian. See photo below . Temperature Display Interior temperature is indicated in degrees C by the solar powered LCD display mounted in the header panel. Temperature Adjustment Instructions- Model 50FREETSA FMS Refrigerator...
PERATION Material Compatibility The interior cabinet of this unit is constructed of anodized aluminum. Care must be exercised when determining which chemicals may be stored in the refrigerator and freezer sections and which type of storage materials should be employed. ABS Plastic deteriorates when exposed to, but not limited to, the following: Aliphatic Hydrocarbons;...
Troubleshooting This table is intended to assist in resolving user-correctable Refrigerator problems by relating symptoms to their likely causes. If service beyond the scope of this table is required, contact Customer Service at 1-800-438-4851. Symptom Probable Cause Action Does Not Run Unit Unplugged Plug in Unit Blown fuse or tripped...
Maintenance Cabinet Cleaning Warning The cabinet interior should be cleaned frequently. Any When servicing the unit, disconnect spilled liquid should be wiped off immediately. Stains from the electrical power source. resulting from some spills can be permanent if not quickly removed.
(ii) misuse (including use inconsistent with written operating instructions for the product), mishandling, contamination, overheating, modification or alteration of the product by any customer or third party or (iii) use of replacement parts that are obtained from a party who is not an authorized dealer of Thermo Scientific prod- ucts.
Need help?
Do you have a question about the 50FREETSA and is the answer not in the manual?
Questions and answers