3.3.2

Before transport: Attaching the transport base plate

To reattach the base plate, place the hook-and-loop tape underneath the base
plate.
Lift the device onto the base plate so that the ribs on the transport base plate fit
into the slits on the bottom of the device (F
Fasten the hook-and-loop tape around the transport handle.
NOTICE
► Always attach the transport base plate when repackaging the device for trans-
port.
3.4

Installation Instructions

3.4.1

Handpiece components

1
1
Handpiece handle
2
Projectile
3
Allen key (see Cleaning / Overhaul)
4
Guide tube
5
Shaft
6
Open-end spanner
7
Pulse transmitter insert
8
Pulse transmitter screw caps
3.4.2

Assembling the handpiece

CAUTION
If the shaft of the handpiece is damaged or uneven, then when removing or screwing
in the shaft without a suitable tool,
minor scratching injuries may occur.
► Be sure to use the open-end spanner provided to remove or screw in the shaft.
19 800 0002 12.2020
. 3-3 ).
IG
2
4
3
7
5
8
6
14